BREAKDOWN VOLTAGE TESTER

IEC 60156
  • 0-100 KV oil test set in accordance with IEC 60156 with automatic voltage control.
  • Output Voltage : 50 – 0 – 50 KV. i.e. 100 Kv across sample.   (Voltage rise Programmable)
  • Capacity : 1 KVA.
  • Control : Motorized.
  • Metering : Digital.
  • Protection : A fast acting D.C. Relay will sense the primary current of H.T. Transformer and protect the unit.
  • Stirring Facility: Stirrer is provided to stir the oil as per requirement of testing methods.
  • Test Cell : Corona free acrylic type with electrodes as per fig ‘2’ of IEC 60156 but slightly modified in volume to suit the condition of high voltage.
  • Interlocking : Zero start interlocking , H.V. Chamber Interlocking .
  • Gauge : “Go” “NOGO” Gauge of 2.51 mm & 2.49 mm. For adjusting gap.
  • Auxiliaries :The unit will have main switch, fuse, indicating lamps, push button,
  • Input Supply : 230 ± 10%Volt AC, 50 Hz, 1 Phase.